China Industrial Metering Equipment Online Market
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Quick Detail:
microcomputer design, LCD
transmitter enclosure, wall mount and other forms
compact form factor, integrated device integration
precision AD conversion technology, high accuracy, good repeatability
two-point calibration mode, the sensor error correctable
long-term stable and reliable, easy maintenance
precision automatic temperature compensation system
Unlike traditional divide is not sensitive to the dirt
variety of signal output, alarm setting, typical dosing control
diverse installation and maintenance easy.
Description:
Online ph meter / ph tester ph-5520 is an industrial online ph analysis and ph monitoring instrument. The online ph meter /ph tester is with various software functions, and has such features as easy installation and operation, strong anti-interference capability and little amount of maintenance work. High-accuracy temperature compensation and pH operation compensation is also adopted for this ph meter.
The online ph meter /ph tester is Widely used for online pH monitoring in environmental protection water treatment, pure water treatment, industrial process, agricultural industry etc.
Applications:
1. Chemical Industry: Pharmaceuticals, chlor-alkali chemical, pigment dyes, pulp and paper, intermediates, fertilizers, starch, cornstarch, pesticides, metallurgy, adhesives, detergents and other industries.
2. Water: air scrubbers, boiler feed water, cooling water, heat and power plant water use;
3. Wastewater: wastewater, industrial wastewater, heavy precipitation, sludge dewatering
Specifications:
Model |
5500 |
5510 |
5520 |
|
Power supply |
DC24V |
AC110 |
AC220V |
|
Supply frequency |
---------- |
50/60 |
50/60 |
|
Measurement range |
pH |
0.00~ 14.00 |
||
ORP |
-1000mV~ +1000mV |
|||
Temp. |
0.0 ~99.9°C |
|||
Resolution |
pH |
0.01 |
||
ORP |
1mV |
|||
Temp |
0.1°C |
|||
Indication error |
pH |
≤±0.05 |
||
ORP |
≤±5 mV |
|||
Temp |
≤ ±0.5°C |
|||
Input impedance cause value error |
pH |
±0.03 |
||
Approx equivalent input impedance |
5×1012Ω |
|||
PH buffer solution |
10.0;9.18;7.00;6.86;4.00; 4.01 |
|||
Temp. compensation error |
pH |
≤±0.05 |
||
Entire instrument error |
pH |
≤±0.1 |
||
Entire instrument repeatability error |
pH |
0.05 |
||
ORP |
±2mV |
|||
Compensation range |
Temp. |
0 ~ 50°C |
||
Working environment |
Temp. |
0 ~ 50°C |
||
Humidity |
≤85% RH |
|||
Cable length |
≤20m(standard 10m) |
|||
Control output |
double relay(single contact ON/OFF),hysteresis can be set |
|||
Contact capacity |
3A(max), AC220V/AC110V 3A(max), DC24V
|
|||
Power consumption |
<2.5W |
|||
(4~20)mA current loop
|
Channels |
One channel |
||
Technical features |
Related Search
Measuring Instruments
Analysis System
PH Meter
PH Tester
Measurement Instrument
Measurement Meter
More>>