Voltage Meter / Multimeter Electrical Calibrators With 8 Inch Color Touch Screen
Main function:
AT1013 can calibrate all kinds of Multimeter below Class 0.5, Voltage meter, Ammeter, Ohmmeter,
and so on.
Main features:
8 inch color touch screen, interface friendly, easy to operate;
The outputs of voltage, current and resistance can be adjusted continuously;
Applied as AC&DC standard sources;
Equipped with RS-232, AT1013 can support both stand-alone operation and PC control;
Self-protection, alarming and displaying fault location for output overload, Voltage short circuit and
Current open-circuit;
Automatically detect, diagnose and alarm for failure.
Remotely updating online, easily achieve software update;
With multifunction, the equipment can calibrate Multimeter, Voltage meter, Ammeter, Ohmmeter;
Output 2nd ~63rd harmonics;
Support calibrating by userself.
Technical specification:
AC Voltage output:
Range: 2V, 20V, 200V, 700V
Adjusting range: (0~120) %RG, RG refers to range, similarly hereafter
Adjusting resolution: 0.01%RG, 0.1%RG, 1%RG, or 10% RG
Stability: 0.02%/2min
Distortion: ≤0.2% (Non capacitive load)
Max output load: 25V and upper ranges: 20VA each phase
Accuracy: 0.1%RG
AC Current output:
Range: 2mA, 20mA, 200mA, 2A, 25A;
Adjusting range: (0~120) %RG, RG refers to range, similarly hereafter;
Adjusting resolution: 0.01%RG, 0.1%RG, 1%RG, or 10% RG
Stability: 0.02%/2min
Distortion: ≤0.2% (Non capacitive load);
Max output load: 20VA (20A range);
Accuracy: 0.1%RG
Frequency Output
Adjusting range: 50Hz, 400Hz;
Harmonic setting
Harmonic order: 2nd ~63rd ;
Harmonic amplitude: 0~40%;
Harmonic angle: 0°~359.99°;
Harmonic set error: 2~31times: ≤±0.1%,32~63 times: ≤±0.2%
DC Voltage output
Range (4 wires): 200mV, 2V;
Range (2 wires): 20V, 200V, 750V, 1000V;
Adjusting range: (0~120) %RG; 1000V: (0~100) %RG
Adjusting resolution: 0.01%RG, 0.1%RG, 1%RG, or 10% RG;
Stability: 0.01%/2min (Class 0.05), 0.02%/2min (Class 0.1)
Output load: ≥50mA;
Accuracy: 0.1%RG;
Ripple content: no more than 1%.
DC Current output
Range: 200μA, 2mA, 20mA, 200mA,2A, 25A;
Adjusting range: (0~120) %RG;
Adjusting resolution: 0.01%RG, 0.1%RG, 1%RG, or 10% RG;